The PTX 7500 Series of current output sensors use a
modular construction to offer customers a wide choice
of configurations at an economic cost. This is combined
with fast delivery to meet the demands of today's
market.
Operating Pressure Ranges
Any zero based full scale (FS) from 1.5 to 1000 psi (100
mbar to 70 bar) gauge or absolute
Any zero based FS above 1000 psi (70 bar) up to 10000
psi (700 bar) sealed gauge or absolute
Barometric 11.5 to 17.4 psi (800 to 1200 mbar) absolute
For other barometric, elevated zero (e.g. 15 to 60 psi
(1 to 4bar)) and compound ranges (e.g. -15 to +15 psi
(-1 to +1 bar)) Refer to GE Druck for further information
Pressure Units
psi, mbar, bar, hPa, kPa, MPa, mmH2O, torr, kgf/cm2
cmH2O, mH2O, inH2O, ftH2O, mmHg, inHg, kg/cm2
Pressure Containment
The operating FS pressure range may be
exceeded by the following multiples without a loss of
mechanical containment:-
Gauge ranges:
• 12 x for ranges up to 2.5 psi (160 mbar)
• 8 x for ranges above 2.5 to 7.5 psi (160 up to 500 mbar)
• 6 x for ranges above 7.5 to 30 psi (500 mbar up to 2 bar)
• 4 x for ranges above 30 to 1015 psi (2 up to 70 bar)
(2900 psi (200 bar) max)
Sealed gauge and absolute ranges:
• 3625 psi (250 bar) for ranges 1.5 to 2030 psi (100 mbar
to 140 bar)
• 14500 psi (1000 bar) for ranges above 2030 up to
10155 psi (140 up to 700 bar)
Pressure Equipment Directive (PED) 97/23/EC
approved (Category 1 - Pressure Accessory)
Performance
Accuracy
Combined effects of non-linearity, hysteresis and
repeatability
±0.1% FS Best Straight Line (BSL) typical (±0.2% FS BSL
max)
0.15% Terminal Straight Line (TSL) typical (0.3% TSL max.)
Zero Offset and Span Setting
Supplied with ±5% zero and span noninteractive site
adjustable potentiometers (excludes 7533 model)
Long Term Stability
±0.1% FS per year
Operating Temperature Range
-40 to 212°F (-40 to 100°C) ambient (176°F (80°C) max for
7511/7533/7534 models)
-40 to 248°F (-40 to 120°C) process media
Temperature Effects
Output will not deviate from room temperature
(RTE) by more than:
• 0.7% FS typical (1% FS max) over 14 to 122°F (-10 to 50°C)
• 1.5% FS typical (2% FS max) over -4 to 176°F (-20 to 80°C)
For ranges below 500 bar, these values will multiply on
a pro rata basis.
Shock
• 1000g, half sine pulse, duration 1 msec
• 100g, peak half sine wave duration 11 msec
• 2000g, half sine pulse, duration 0.5 msec
All in each of 3 mutually respendicular axis will not effect
calibration.
Vibration
Conforms to MIL-STD 810C method 514.2 figure 514.2-2
curve L
Pressure Response
1kHz band width (63% response to step change in
pressure).